Resources

Philips Zymed Holter Error Codes 501,201, 202, 204, 214, 251 and 257

The following error codes can occasionally be displayed on the LCD display on the DigiTrak XT 48 Hour Holter monitors.

 

Error Message

Error: ### – Configuration read failure, where ### is between 201 and 202, 204 and 214, and 251 and 257 inclusive.

Causes and/or Message location

The firmware has encountered a problem with reading the external configuration file.

Resolution

– Reload the recorder’s upgrade file.

– If problem persists, contact Philips Customer Service.

 

 

Error Message

Error: 501 – External flash failure. – The firmware has encountered a problem trying to access the external flash.

Causes and/or Message location

Message(s) can be displayed at various points during recorder operation.

Resolution

– Remove and reinsert the batteries.

– If problem persists, contact Philips Customer Service.

Share